The Batteries Discharge Very Quickly, What Was Done Wrong?
// November 12th, 2012 // Uncategorized
There are two typical mistakes leading to a quick discharge of batteries:
- the Toggle switch at the bottom left corner of the Smart Tweezers unit is returned to the default position after voltage measurement is done; and
- the Test Frequency is set to 10 KHZ. In the latter case, Smart Tweezers do not turn off automatically due to parasitic capacitance of the tweezers’ tips leading to a current flowing through it. Change the Test Frequency to ‘Auto’ after measuring something at high frequency
Leave a Reply