The Batteries Discharge Very Quickly, What Was Done Wrong?

// November 12th, 2012 // Uncategorized

There are two typical mistakes leading to a quick discharge of batteries:

  1. the Toggle switch at the bottom left corner of the Smart Tweezers unit is returned to the default position after voltage measurement is done; and
  2. the Test Frequency is set to 10 KHZ. In the latter case, Smart Tweezers do not turn off automatically due to parasitic capacitance of the tweezers’ tips leading to a current flowing through it. Change the Test Frequency to ‘Auto’ after measuring something at high frequency